Physical:

  • CASE: 9.8" X 3.9" X 2.9"Aluminum case. Weight: 3 lbs.
  • PROBE: 9"L. X 3/8" OD. Inconel stack probe. Probe housing connects to instrument via a 10 ft. Viton(R) hose and water trap with fiber filter. Max. continuous temperature: 2000 deg. F.

Viton(R) is a registered Trademark of DuPont Dow Elastomers.

Electrical Power:

  • BATTERY: 6VDC. Interchangeable rechargeable NiCd or four disposable AA alkaline cells. 6 approximately hours operating time.
  • AC: 120V. 60 Hz. std. (220V. 50 Hz. optional), using battery charger (NiCd batteries only!)

 

Display:

Four line by 16 character wide temperature range LCD with backlight illumination.

Measured Parameters

Range

Resolution

Accuracy

1. AMBIENT TEMPERATURE
IC sensor. Degrees F or C.
0-150 degrees F 1 degree F or C. 3 degrees F.
2. STACK TEMPERATURE
Type K thermocouple.
Degrees F or C.
0-2000 degrees F (1100 C). 1 degree F.(1 C.) 5 degrees F.
3. OXYGEN
Electrochemical cell.
Life 2 years.
0-25% 0.1% 0.2%
4. NITRIC OXIDE (NO).
Electrochemical cell.
Life 2 years.
Other ranges are available.
0-2000 PPM. 1 PPM 4% of reading.
5. NITROGEN DIOXIDE (NO2).
Electrochemical cell.
Life 2 years.
Other ranges are available.
0-1000 PPM 1 PPM 4% of reading
6. CARBON MONOXIDE (CO)
Electrochemical cell.
Life 2 years.
Other ranges are available.
0-2000 PPM 1 PPM 4% of reading
7. SULFUR DIOXIDE
Electrochemical cell.
Life 2 years.
Other ranges are available.
0-2000 PPM 1 PPM 4% of reading
8. COMBUSTILBES
0-5% 0.1% 10% (CH4)
9. STACK DRAFT
+10" to -40" WC .1" 5% of reading
10. TIME/DATE
Time in hours, minutes, seconds. Date in month, day, year format.

 

 

Computed Parameters

Range

Resolution

Accuracy

1. COMBUSTION EFFICIENCY
Heat Loss Method
0-100% 0.1% 1%
2. CARBON DIOXIDE (CO2)
0-40% 0.1% 5% of reading
3. EXCESS AIR
0-1000% 1% 10% of reading
4. OXIDES OF NITROGEN (NOx)
0-3000 PPM 1 PPM 4% of reading

Printer:

External ENERAC 2", 28 char. per line thermal printer.

Storage:

  • INTERNAL: 400 individually selectable buffers hold one complete set of measurements each in non volatile memory. Buffer contents can be sent to printer or RS-232 port.
  • EXTERNAL:Data storage performed either on command or on a periodic basis.


Communications:

  • RS-232 PORT. RS-232C port (DTE), 9600 baud, half duplex, 1 start bit, 8 data bits, 1 stop bit, no parity.
  • USB PORT. USB Type B Port. Driver included.
  • SOFTWARE Over 20 software commands for diagnosis and measurement. Windows software available.
  • Bluetooth


Miscellaneous:

  • FUELS: 11 fuels, Custom fuels available on request or by customer programming using Enercom(TM) software..
  • CALIBRATION: Optional Autozero. Automatic software span calibration CO, NO, NO2 and SO2.