Main Page
Model
3000E
NITRIC OXIDE(NO) |
Electrochemical (SEM) Cell. Life: Approximately 3 years.
Accuracy: 2% of reading (*)
(*) When tested according to 40 CFR 60, RAA test | ||
NITROGEN DIOXIDE(NO2) |
Electrochemical (SEM) Cell Life: Approximately 3 years.
Accuracy: 2% of reading | ||
CARBON MONOXIDE(CO) |
Electrochemical (SEM) Cell. Life: Approximately 3 years.
Accuracy: 2% of reading (*) (*) When tested according to 40 CFR 60, RAA test | ||
SULFUR DIOXIDE(SO2) |
Electrochemical (SEM) Cell. Life: Approximately 3 years.
Accuracy: 2% of reading (*) (*) When tested according to 40 CFR 60, RAA test. | ||
COMBUSTIBLES |
Catalytic sensor Life: indefinite. Range: 0-6.00% Resolution: 0.01% Accuracy: 10% of reading in CH4 gas | ||
VELOCITY(Optional) |
S type pitot tube. Range: 0-200 ft/sec. (2" W.C.) Resolution: 1 ft/sec Accuracy: Meets EPA Method 2 | ||
|
Electrochemical cell. Life: Approximately 3 years. Range: 0-25% Resolution: 0.1% Accuracy: 0.2% | ||
TEMPERATURE |
IC sensor. Degrees F or C. Range: 0-150 degrees F Resolution: 1 degree F or C Accuracy: 3 degrees F | ||
TEMPERATURE |
Type K thermocouple. Degrees F or C Range: 0-2000 degrees F (1100 C). Resolution: 1 degree F.(1 C.) Accuracy: 5 degrees F. | ||
| Time in hours, minutes, seconds. Date in month, day, year format. | ||
ENERAC'S PROPRIETARY LED BASED | |||
|
Warm-up Time: 60 seconds Operating Humidity: 0-90% | |||
CARBON MONOXIDE (NDIR) |
Range: 0.01 % to 15.0 % Resolution: 0.1 % Accuracy: 5 % of Reading | ||
CARBON DIOXIDE (NDIR) |
Range: 0 - 20 % Resolution: 0.1 % Accuracy: 5 % of Reading | ||
HYDROCARBONS (NDIR) | Range: 0- 30,000 PPM Resolution: 1 PPM Accuracy: 0 - 2,000 PPM 3% of Reading 2,000 - 15,000 PPM 5% of Reading 15,000 - 30,000 PPM 8% of Reading | ||
EFFICIENCY |
Heat loss method. Unique four loss factor computation. (dry gas, water
vapor, gaseous combustibles, combustibles in ash) Range: 0-100% Resolution: 0.1% Accuracy(4 loss): 1% (above H2O condensation) 2% (below H2O condensation) (**) (**) For use with ultra high efficiency burners that operate below the condensation point. | ||
CARBON DIOXIDE(CO2) |
Range: 0-40% Resolution: 0.1% Accuracy: 5% of reading. | ||
|
Range: 0-1000% Resolution: 1% Accuracy: 10% of reading | ||
OXIDES OF NITROGEN(NOx) |
Accuracy: 2% of reading (*) | ||
|
Range: 0-2500 milligrams/cubic meter (CO, NO, NO2, NOX, SO2) Resolution: 2 mg/M3 Accuracy: 5% of reading (Oxygen correction factor for ppm and milligrams adjustable 0-20% in 1% steps plus TRUE) | ||
|
Range: 00.00-99.99 lbs./million BTU (CO, NO, NO2, NOX, SO2) Resolution: 0.01 lbs./MMBTU Accuracy: 5% of reading | ||
|
Range: 00.00-99.99 grams/brake hp-hr (CO, NO, NO2, NOX, SO2) Resolution: 0.01 arms/bhp-hr Accuracy: 10% of reading | ||
|
Range: 0-99.99 lbs./hour (CO, NO, NO2, NOX, SO2) Resolution: 0.01 lbs./hour Accuracy: 10% of reading Range: 0-99.99 tons/day for (CO2) | ||
STACK GAS FLOW RATE(Optional) |
Range: 0-65000 CFM Resolution: 1 cfm Accuracy: Meets EPA Method 2 |
|
For Proper NO2 and SO2 Readings, the Standard Emissions Probe contains a
battery operated Permeation Drier (40 CFR 60, Appendix A Method 6C Sec.
5.1.5) 24"L. X 3/8" OD. Inconel probe with Hastelloy X sintered filter and ½" deflector mounted on permeation drier housing. Probe housing connects to instrument via a 10 ft. Viton(R) hose. Max. continuous temperature: 2000 deg. F. Max. sample dew point (past drier): 50 deg. F.@ 600 cc/min. (Natural gas fuel @ 0% oxygen). For lower dew points an optional supplemental moisture removal is available.
| VELOCITY-EMISSIONS PROBE (VEP) ASSEMBLY
|
To meet the Performance Requirements of EPA's Method 2, the VEP option
is available. Three part 316 stainless steel detachable welded assembly. Consists of 3/8" OD sample probe with sintered filter, two sections of 3/16" S type pitot tube and sheathed type K thermocouple. Standard length 17". (Specify desired size of optional length). Max. Continuous temperature: 1700 deg. F. |
| 9 Pin - RS-232c port (switch selectable DTE or DCE ), 300-9600 (1200 default) baud, user selectable, half duplex, 1 start bit, 8 data bits, 1 stop bit, no parity. |
| Internal 1200 baud modem connects to a modular phone line for remote communication. (A 50 word command set sends instructions to the unit for remote control and troubleshooting) |
| Eight analog outputs, 0-5 VDC, of the following parameters: stack temperature, ambient temperature, oxygen, combustibles, carbon monoxide, nitric oxide, nitrogen dioxide and sulfur dioxide. |
| ENERCOM for Windows(TM) software. 3 1/2" diskette includes MONITOR, ALARMS, FUEL PROGRAMMING, BAR GRAPHS and LINE PLOTS including CUMULATIVE PLOTS of mass emissions. It calculates emission averages, Standard Deviations, identifies Min and Max values. All data is saved in a variety of common formats for use directly in commercially available spreadsheets. |
CASE: | 18"X13"X6" Aluminum carrying case with lock. Weight: 24 lbs | ||||||||||
DISPLAY: | 0.35" high by 24 Character two line LCD with backlight illumination and adjustable viewing angle. | ||||||||||
PRINTER: |
SEIKO 4",40 char. per line thermal printer with line feed button and with end of paper override.
Operates in any of four print modes: 1. TEXT MODE. 25 line printout of instant. values of all measured parameters. (time req.20 sec.) 2. PLOT MODE. Any one parameter vs. time plotted. 3 ordinate scales: full, half, quarter. Time scale: Selectable, 1 sec/dot-1 min/dot in 1 sec/dot intervals. 4 parameter plot CO-OXYGEN-NOX-EFFICIENCY, single scale. | ||||||||||
CALIBRATION CERTIFICATION PROTOCOL | Automatic printout of both auto zero and span calibration test results, including sensor diagnostics and filter performance status. The protocol provides the user with an early warning of sensor and/or filter deterioration or failure. | ||||||||||
EXTERNAL PRINT MODE | Prints messages sent via RS-232 port. | ||||||||||
STORAGE: | 1. INTERNAL 50 individually selectable buffers each of which
holds one complete set of measurements in non volatile memory. Buffer
contents can be sent to printer or RS-232 port 2. EXTERNAL (OPTIONAL) See Enercom software. | ||||||||||
ELECTRICAL POWER OPTIONS |
| ||||||||||
MISCELLANEOUS |
|
NO SENSOR TEMPERATURE CONTROL
(Automatic) To avoid sensor zero drift, which is a function of the interaction
between the temperature (Arrhenius Law), and the NO sensor's previous exposure
history, the temperature of the NO sensor is nominally kept at 25 Degrees Celsius
(See CTM-022 Secs. 2.3.1.
and 7.2.1.)
CALIBRATION CERTIFICATION PROTOCOL
(Automatic) Automatic printout of both auto zero and span calibration test results, including sensor diagnostics and filter performance status. The protocol provides the user with an early warning of sensor and/or filter deterioration or failure. (See CTM-022 sec. 6.6.)
Note:
The Performance Specifications are Equivalent to EPA's traditional NOx
Test Method - 40 CFR 60, Appendix A, Method 7E.
CTM-022 Section 4.0 Measurement System Performance Specifications:
| 4.1 | Measurement System Calibration Error | +/- 2% of High Range Calibration Gas |
| 4.2 | Sampling System Bias Test | +/- 5% of High Range Calibration Gas |
| 4.3 | Zero Drift | +/- 3% of High Range Calibration Gas |
| 4.4 | Calibration Drift | +/- 3% of High Range Calibration Gas |
Enerac