NONDISPERSIVE INFRARED (NDIR) SENSOR BENCH
Measurement of CO - CO2 - Hydrocarbons by infrared technology

The new Enerac
700
IR and
3000E SEM-IR measures hydrocarbon concentrations using the latest state-of-the-art
infrared technology.
The new hydrocarbons measurement capability replaces the previous combustibles measurement that used catalytic technology and is offered as an add-on option for customers.
The hydrocarbon measurement features:
- Meets EPA's Reference Method 25B "Determination of Total Gaseous Organic Concentration Using a Nondipersive Infared Analyzer"
- Measures Hydrocarbons as propane using the latest state of the art pulsed infrared LED emitter and dual lead selenide detector technology
- Exceptional stability and accuracy at very low power consumption (less than 3 watts)
ENERAC NDIR Specifications
|
Measurement gas
|
Range
|
Resolution
|
Accuracy
|
Hydrocarbons
(as propane)
|
0-30,000 PPM (0-3%) by volume
|
1 PPM
|
+/-3% of reading
|
Carbon Monoxide
(CO)
|
0-15%
|
0.01 %
|
+/-3% of reading
|
Carbon Dioxide
(CO2)
|
0-20%
|
0.1 %
|
+/-3% of reading
|
Response Time: <10 sec
Warm up time (@22 deg. C): 60 seconds
Operating Temperature range: -10 to 40 deg. C
Operating humidity: 5-95% non condensing
|